Specifications
|
|
| Application |
- Non-destructive leak detection for packages with defect profile typically >15 microns
- Altitude package testing
|
| Package Type |
- Filled & sealed pouches, cups, trays
- Empty trays
- Tyvek® lidded blisters
- Pouches
|
| Package Materials & Combinations |
Tyvek®, Paper, Foil, Film, Aluminum, Plastic & Laminated Materials |
| Offline or Online |
Offline lab instrument |
| Test Configuration |
- Offline laboratory
- Production line applications
|
| Test System |
Single Transducer PERMA-Vac Technology* |
| Test Method* |
Vacuum Decay |
| Operator Interface |
10" color touch screen |
| Test Parameter Storage |
Up to 20 packages |
| Test Sensitivity |
2 ccm (approximate hole size 15 microns) |
| Test Results/Resolution |
Pass/Fail Result in mBar units |
| Security Password |
Yes |
| Remote Internet Access |
Yes |
| Data Collection |
View on touch screen and electronic data log collection |
| Test Chamber |
Manual or semi-automatic |
| ASTM Test Method |
ASTM F2338-09** and ASTM D6653-13 |
| Test Instrument Enclosure |
Stainless steel |
| Test Dimensions |
14.5” W – 22” D – 12” H |
| Weight |
35 lbs |
| Power |
100-240 VAC; 50/60 Cycles |
| Air |
90 psi |
| Options |
- Validation Qualification Package (IQ/OQ/PQ)
- Microcalibrator Flowmeter
|
* U.S. Patents 5,513,516 6,513,366 8,544,315.
** ASTM Test Method F2338-09 - www.astm.org based on VeriPac leak testers. |
| Test results may vary according to application and package specifications. |