Why Vacuum Decay Is Suitable for Both Dry and Liquid-Filled Packages?

Why Vacuum Decay Is Suitable for Both Dry and Liquid-Filled Packages?

A batch of sterile injectables cleared your QC line on a Tuesday morning. By Thursday, a single compromised vial had failed container closure, and the contamination was traced back to a pinhole defect smaller than 20 microns. The package looked perfect. The seal appeared intact. Your dye ingress test gave you a clean result. That scenario plays out more often than the industry likes to admit, and it costs manufacturers in recalls, regulatory actions, and far worse: patient safety events.

The question isn't whether package integrity failures happen. They do. The real question is whether your testing method catches them before they leave your facility, and whether it can do so without destroying the sample in the process.

That's where Vacuum decay testing standardized under ASTM F2338, has earned its place as one of the most reliable container closure integrity testing (CCIT) methods available today. And unlike many other techniques, it works with equal precision on dry-product packages and liquid-filled packages, without modification, without solvents, and without sacrifice of the sample.

What does ASTM F2338 Measure?

At its core, vacuum decay is a pressure-based leak detection method. The test places a package inside a sealed, evacuated test chamber. The chamber is brought to a defined vacuum level, typically in the range of 1 to 5 mbar absolute. Once that vacuum is established and stabilized, the instrument monitors the chamber pressure for a defined dwell time, usually between 5 and 30 seconds.

If the package is intact, the chamber pressure holds steady. If there's a breach, a microhole, a compromised seal, a crack in a rigid container, gas or vapor escapes from the package into the chamber. That escape registers as a measurable pressure rise. The magnitude and rate of that pressure rise directly correlate to leak size and location.

Key principle: ASTM F2338 does not rely on detecting liquid, dye, or tracer gas. It detects pressure change — a universal physical phenomenon that occurs regardless of what is inside the package.

Dry Filled Packages vs. Liquid Filled Packages: A Side-by-side Comparison

Parameter Dry-filled packages Liquid-filled packages Common ground
Detection mechanism Gas flow (viscous and molecular) Vapor / dissolved gas escape at reduced pressure Pressure rise in evacuated chamber
Primary physics Knudsen / Poiseuille flow Henry's Law, vapor pressure, degassing Pressure differential across defect
Vacuum depth Lower vacuum often sufficient Requires careful optimization to avoid false signals Defined per ASTM F2338 validation
Key risk in method dev Headspace composition variability Over-evacuation causing false positives Requires positive and negative controls
Typical sensitivity 2–10 micron defects 5–20 micron defects (liquid-dependent) Deterministic, quantitative output
Best application Blisters, sachets, lyo vials, pouches Parenterals, prefilled syringes, ampoules Any hermetically sealed primary container

Advantages that matter on the plant floor

  • Non-destructive by design. The package is not opened, punctured, submerged, or exposed to reagents. It can be returned to the batch after testing, which is critical for high-value biologics and small-volume parenterals where destructive testing means discarding saleable product.
  • Quantitative, not interpretive. Unlike dye ingress or bubble emission, vacuum decay gives a numerical pressure-rise value. Pass/fail decisions are based on thresholds, not technician judgment. That's important for audit trails and regulatory submissions.
  • Rapid cycle times. Most tests complete in 30–90 seconds per unit. High-throughput configurations can test multiple packages simultaneously, making 100% testing, rather than AQL sampling, a realistic option for critical products.
  • Repeatable and reproducible. Because the method is instrument-driven, inter-operator and inter-laboratory variability is significantly lower than visual or dye-based methods. This supports validation under ICH Q2(R1) requirements.
  • No consumables or hazardous materials. Unlike Helium Leak detection, there's no tracer gas required. Unlike dye ingress, there's no methylene blue or other colorant. Operational costs are low and disposal considerations are minimal.
  • Works across container types and sizes. From 0.5 mL lyophilized vials to 500 mL IV bags, instrument manufacturers offer chamber configurations that accommodate a wide range of primary containers. One method, multiple formats.

How Vacuum Decay Fits within Broader CCIT Strategy?

In the context of CCI technologies, vacuum decay occupies the middle ground between probabilistic methods (dye ingress, bubble emission) and high-sensitivity tracer methods (helium mass spectrometry, headspace analysis). It offers better sensitivity than dye ingress and far lower equipment and operational costs than helium-based systems.

For many manufacturers, vacuum decay becomes the primary release test method, with headspace analysis or helium leak testing reserved for development-stage defect characterization. This tiered approach, endorsed in USP <1207>, allows facilities to optimize both sensitivity and throughput at each stage of the product lifecycle.

Regulatory agencies, including the FDA and EMA, have increasingly emphasized deterministic CCI methods over probabilistic ones, particularly for sterile products. Vacuum decay, validated per ASTM F2338, qualifies as a deterministic method — a distinction that carries weight in both NDA submissions and inspection responses.

Frequently Asked Questions

1. What defect sizes can vacuum decay detect in liquid-filled vials?

For aqueous liquid-filled vials, validated vacuum decay methods typically achieve detection of defects in the 5–20 micron range, depending on the liquid formulation, headspace volume, and test parameters. Viscous or high-density formulations may reduce sensitivity slightly. Specific detection limits must be established during method validation using calibrated positive-control samples with known defect sizes.

2. Is vacuum decay per ASTM F2338 accepted by the FDA for container closure integrity testing?

Yes. The FDA recognizes ASTM F2338 as a validated standard for vacuum decay leak testing. The method is classified as a deterministic CCIT technique under FDA's 2008 guidance on container closure systems and aligns with the framework described in USP <1207>. Manufacturers are expected to validate the method for their specific container-closure system and product type.

3. Can vacuum decay be used for 100% inspection rather than sampling?

Yes, and this is one of its key operational advantages. With cycle times of 30–90 seconds per unit and multi-chamber instrument configurations, 100% inline or at-line testing is achievable for many production formats. This eliminates the statistical uncertainty of AQL-based sampling, which is particularly valuable for sterile injectables and other high-risk product categories.

4. How does vacuum decay compare to dye ingress testing for regulatory purposes?

Dye ingress is classified as a probabilistic method — its reliability depends on operator technique, dye concentration, immersion time, and visual inspection variability. Regulatory guidance increasingly favors deterministic methods like vacuum decay for product release testing, particularly in the sterile injectable space. Dye ingress may still be useful as a development tool or for method comparison during validation, but it is not recommended as a standalone release test for parenteral products.

5. What validation studies are required to qualify a vacuum decay method under ASTM F2338?

A complete validation package typically includes: instrument qualification (IQ/OQ/PQ), method development studies to optimize test parameters (vacuum level, stabilization time, test time), sensitivity determination using positive-control samples with calibrated defects, specificity studies to confirm the method does not generate false positives from normal package variation, and reproducibility and repeatability studies across operators, days, and instruments. Reference standards from PTI, Lighthouse Instruments, or equivalent suppliers provide calibrated leak standards for this purpose.

Readmore...
cci technologies, vacuum decay leak test, ASTM F2338, container closure integrity testing, container closure integrity, package integrity testing
6
22
May 2026

The Impact of Low Headspace on Leak Detection Accuracy

The Impact of Low Headspace on Leak Detection Accuracy?

Pharmaceutical packaging has grown considerably more complex over the past two decades. Prefilled syringes, blow-fill-seal (BFS) containers, cartridges for autoinjectors, and highly concentrated biologic formulations have all become standard parts of the product landscape. What many of these formats share is minimal internal headspace, and that matters more than most people outside the CCIT world appreciate.

Headspace is not just an incidental feature of container design. For most pressure-based and gas-based leak detection technologies, it is the measurement medium. When headspace volume drops, so does the detectable signal that a leak generates. The result is a narrower, noisier measurement window that demands both better equipment and more rigorous method development.

Container closure integrity testing (CCIT) for low-headspace products is not simply a matter of applying a standard method and tightening the acceptance criteria. It requires a fundamentally different approach to test configuration, sensitivity optimization, and validation strategy. Getting it wrong has real consequences, from compromised sterile products reaching patients to expensive false rejects disrupting manufacturing.

What are the Pharmaceutical Formats Commonly Affected by Low Headspace?

Pharmaceutical formats commonly affected by low headspace include:

  • Prefilled syringes: Typically filled near nominal volume with minimal residual gas space.
  • BFS (Blow-Fill-Seal) containers: Often sealed with near-zero headspace depending on fill design.
  • Cartridges for autoinjectors and pen injectors: Compact formats with tightly controlled fill volumes.
  • Small-volume vials (1–2 mL): Limited internal volume by design.
  • Biologic and cell therapy packaging: Often filled under modified atmosphere with controlled, minimal headspace.
  • Ophthalmic unit-dose containers: Single-use BFS or thermoform formats with near-complete fill.

The role of headspace in leak detection is straightforward but easily underestimated. When a defect is present, gas either enters or escapes the container through the leak pathway. The pressure change or tracer gas signal that results is proportional to the internal volume available to participate in that exchange. Less headspace means a smaller, faster-attenuating signal.

What are the Risks of Inaccurate Leak Detection in Low-Headspace Containers?

When a CCIT method is not well-suited to a low-headspace product, the consequences span both directions of the quality spectrum. Neither outcome is acceptable in sterile pharmaceutical manufacturing.

False Accepts: The Invisible Risk

  • Compromised containers pass testing and enter the supply chain.
  • Microbial ingress through sub-visible defects can compromise sterility without visible product change.
  • Oxygen or moisture entry can degrade potency in biologics, lyophilized drugs, and oxygen-sensitive formulations.
  • Shelf life is shortened, potentially causing out-of-specification results at stability timepoints.
  • Patient safety risk if sterile barrier failure goes undetected.

False Rejects: The Operational Risk

  • Intact, conforming containers are incorrectly rejected, increasing manufacturing loss.
  • High false reject rates can mask real quality trends by creating background noise in the data.
  • Investigation burden increases as QA attempts to explain non-reproducible failures.
  • Distorts process capability data and complicates ongoing validation maintenance.

Why CCI Method Optimization Matters?

In low-headspace applications, using a test method with insufficient sensitivity is not a conservative approach; it is an unreliable one. An undertested product is not a safe product. Deterministic CCIT methods designed and optimized for low-headspace formats are the only reliable path to accurate, defensible test results.

What are the Strategies for Testing Low-Headspace Products?

1. Helium Leak Detection

Why is helium leak detection effective for low-headspace products?

Helium leak detection uses helium as a tracer gas with an extremely small atomic radius and near-zero atmospheric background (~5 ppm). Because detection relies on the presence of helium atoms rather than a bulk pressure change, sensitivity is not directly constrained by headspace volume. This makes it highly effective for low-headspace containers where pressure-based signals are weak.

Helium Leak detection works by filling or purging the container headspace with helium, then placing the package in a vacuum chamber or scanning it with a mass spectrometer sniffer probe. The instrument detects individual helium atoms escaping through any defect. Because the signal is based on tracer gas concentration rather than bulk pressure differential, it is far less dependent on available headspace volume.

Key advantages for low-headspace applications:

  • Sensitivity independent of headspace volume in tracer gas mode.
  • Detection capability down to 10?¹² mbar·L/s under optimized conditions.
  • Fully deterministic and quantitative, produces an actual leak rate value.
  • Directly correlatable to MALL for validation under USP <1207>.
  • Well-suited to rigid formats: vials, cartridges, ampoules, and prefilled syringes.
  • Atmospheric helium background is extremely low, providing an inherently high signal-to-noise environment.

2. Vacuum Decay Technology

How does vacuum decay testing work for low-headspace products?

Vacuum decay testing evacuates a test chamber containing the sealed container to a defined vacuum level. Gas escaping from a defective container causes a measurable pressure rise in the chamber. For low-headspace products, test parameters, including chamber geometry, vacuum level, and stabilization time, must be carefully optimized to maximize the signal-to-noise ratio.

Vacuum decay is a non-destructive, deterministic method that requires no tracer gas or sample preparation. The container is tested in exactly the condition it would be distributed. For standard-headspace products, it is one of the most widely used CCI technologies in pharmaceutical manufacturing.

For low-headspace containers, vacuum decay remains effective when properly optimized:

  • Chamber geometry: a tightly fitting test chamber minimizes dead volume, concentrating the signal from any leak.
  • Vacuum level selection: optimized to maximize differential between headspace pressure and chamber pressure.
  • Extended stabilization and measurement intervals: allow weak signals to accumulate above the noise floor.
  • High-resolution pressure transducers: necessary to detect small pressure changes in low-headspace conditions.
  • Temperature-controlled test environments: reduce thermal artifact pressure signals that can obscure real leak data.

Helium Leak Detection vs. Vacuum Decay: Technology Comparison for Low-Headspace Products

Criterion Helium Leak Detection Vacuum Decay
Operating principle Tracer gas (He atoms) detected by mass spectrometry Pressure rise measurement in evacuated test chamber
Sensitivity Extremely high: ~10?¹² mbar·L/s High: low single-digit micron range when optimized
Headspace volume dependency Low — tracer gas detection independent of bulk pressure Moderate — signal is amplified by chamber geometry optimization
Quantitative output? Yes — leak rate in mbar·L/s Yes — pressure differential data correlatable to leak size
Deterministic? Yes Yes
USP <1207> recognized?  Yes Yes
Best-fit formats  Rigid containers: vials, cartridges, ampoules, PFS Broad: vials, bottles, blisters, pouches, PFS, BFS
Key low-headspace advantage  Signal not limited by headspace volume in tracer mode Chamber geometry optimization compensates for low signal
Primary limitation Requires headspace modification; not always releasable post-test Parameter development required for very low headspace

Why Does Deterministic CCIT Matter for Modern Pharmaceutical Packaging?

The regulatory direction set by USP <1207> is clear: deterministic CCIT methods are the preferred approach for package integrity testing of sterile pharmaceutical products, particularly where risk is high. Probabilistic methods, including dye ingress, bubble emission, and microbial challenge tests used in isolation, remain in some protocols. Still, their inability to reliably detect sub-10-micron defects or generate quantitative data makes them increasingly difficult to justify for high-risk products.

Low-headspace products make this case even more sharply. A dye ingress test applied to a prefilled syringe with 0.3 mL of headspace does not generate defensible leak rate data. It tells you whether dye crossed a barrier under specific immersion conditions. It does not tell you whether a 5-micron defect was present, what the headspace exchange rate was, or whether the package meets its MALL.

FDA inspectors reviewing CCIT programs for parenteral and biologic products increasingly expect manufacturers to justify their chosen method against the product's MALL. For any product where the MALL falls below 10 μm, that justification needs to be rooted in deterministic data. Low-headspace products are not an exception to that expectation, if anything, they make meeting it harder, which is exactly why method selection and optimization matter from the earliest stages of packaging development.

Frequently Asked Questions

1. What is headspace in pharmaceutical packaging?

Headspace is the gas-filled volume inside a sealed pharmaceutical container that is not occupied by the drug product. It consists of air, nitrogen, or a controlled inert atmosphere depending on formulation requirements. In leak detection, headspace volume directly determines the magnitude of the pressure or gas signal generated by a defect.

2. Why does low headspace make leak detection more difficult?

Low headspace reduces the internal gas volume available for pressure exchange when a defect is present. This produces a smaller, faster-decaying signal that is harder to distinguish from background measurement noise. It increases signal-to-noise challenges, validation complexity, and the risk of false accepts or false rejects if the test method is not properly optimized.

3. What is signal-to-noise ratio in CCIT and why does it matter?

Signal-to-noise ratio (SNR) in CCIT describes the relationship between the measurable pressure or tracer gas signal produced by a defect and the background variation inherent in the measurement system. A low SNR, typical in low-headspace testing, makes it harder to reliably distinguish real leaks from system noise, increasing the risk of testing errors.

4. Which CCIT technologies are best suited for low-headspace pharmaceutical containers?

Helium leak detection and vacuum decay are the two deterministic CCIT methods with demonstrated capability for low-headspace products. Helium detection is largely independent of headspace volume in tracer gas mode. Vacuum decay, when properly optimized with close-fitting chamber geometry and high-resolution pressure transducers, is effective across a wide range of headspace volumes.

5. What does USP <1207> require for CCIT in sterile pharmaceutical packaging?

USP <1207> requires that container closure integrity testing methods demonstrate sensitivity at or below the product-specific Maximum Allowable Leakage Limit (MALL). It distinguishes deterministic methods — which produce quantitative, measurable data — from probabilistic methods, and supports deterministic testing as the preferred approach for high-risk sterile products.

Readmore...
ASTM F2338, container closure integrity testing, ccit, container closure integrity, cci testing, vacuum decay leak test
16
08
May 2026

What Is Container Closure Integrity Testing A Practical Guide

What-Is-Container-Closure-Integrity-Testing-A-Practical-Guide

What Is Container Closure Integrity Testing (CCIT)?

Container Closure Integrity Testing (CCIT) is the validated process of verifying that a pharmaceutical package maintains a complete sterile barrier, preventing microbial ingress, gas exchange, and contamination across a drug product's entire shelf life. It is a regulatory and quality requirement for sterile drug products, biologics, and parenteral packaging at every stage from development through commercial batch release.

A single undetected 10-micron defect in a vial septum can allow slow microbial ingress over 18 months, invisible to a dye bath, catastrophic at the patient level. FDA Warning Letters have cited insufficient container closure integrity data in NDA submissions as a direct consequence of relying on probabilistic methods alone.

Blue dye ingress and microbial challenge testing dominated CCI practice for decades. Both answer a binary question, did visible dye enter, or not - and both consume the sample. For modern biologics, cell therapies, and high-risk parenteral, that binary answer is no longer sufficient. Quantifiable, reproducible integrity data is now the regulatory and scientific expectation.

CCIT replaces assumption with measurement

The regulatory mandate: understanding USP <1207>

USP <1207>, formally adopted in 2016, is the primary regulatory framework governing container closure integrity testing(CCIT) in the United States. The EMA and ICH Q10 quality system guidelines align with the same principles, making it the de facto global reference for CCI validation.

Its most consequential position: a clear preference for deterministic, non-destructive testing (NDT) over probabilistic methods. Where a deterministic method is technically feasible, USP <1207> expects it to be the default choice. The chapter is structured across three sections — <1207> (general concepts and method selection), <1207.1> (package integrity test methods), and <1207.2> (package seal quality test methods), and requires a risk-based, package-specific justification for every method selected.

A common misconception is that USP <1207> prescribes a single test method. It does not. It mandates a defensible selection process.

Criterion Deterministic Methods Probabilistic Methods
Test Nature Objective, instrument-based Subjective, observer-dependent
Integrity Preservation Non-destructive Destructive
Output Data Quantitative leak rates (mbar·L/s) Qualitative pass/fail only
USP <1207> Standing Preferred where feasible Acceptable only where deterministic is not feasible

Core deterministic CCIT technologies

Method selection must be driven by package design and product characteristics, not instrument availability. The three CCI technologies below represent the most widely validated deterministic methods in pharmaceutical packaging.

1. Vacuum Decay

Vacuum decay places a sealed package inside a test chamber, applies vacuum, and measures any pressure rise via calibrated differential pressure transducers. Pressure rise indicates a leak. ASTM F2338 and validated package-specific studies show that vacuum decay can detect small defects in rigid, nonporous containers, including holes in the 5-micron range under defined test conditions

Best suited for rigid and semi-rigid containers: vials, bottles, blister packs, and prefilled syringes. It is non-destructive, requires no sample preparation, and supports full automation for at-line or 100% inspection. Test parameter optimization, particularly equilibration time and vacuum level, is the primary driver of sensitivity outcomes. In CCIT method development, application-specific parameter optimization is often necessary because off-the-shelf settings may not suit every package, product, or defect challenge.

2. High Voltage Leak Detection (HVLD)

HVLD passes a high-voltage field across a filled container. An intact non-conductive container wall interrupts the circuit; a breach creates a conductive pathway through the product, registering as a measurable resistance change. Under validated conditions, the method detects defects in the 2–10 micron range in liquid-filled parenteral containers.

Its key operational advantage: 100% inline inspection at commercial filling speeds, with no throughput impact. Standard HVLD systems operate at 10–25 kV, a range shown to induce structural changes in low-conductivity, high-concentration biologics including monoclonal antibodies and peptides.

PTI's HVLD reduces applied voltage by approximately 50% while maintaining equivalent detection sensitivity through optimized signal processing. For manufacturers working with biologics above 50 mg/mL in prefilled syringes or vials, this is not a marginal distinction. It directly determines method suitability and product quality risk at the validation stage.

3. Helium Leak Detection

Helium leak detection uses helium as a tracer gas, detected via mass spectrometry. With an atomic radius of 31 pm, helium permeates defects that pressure-based methods cannot resolve, achieving detection limits in the 10?? mbar·L/s range under optimized conditions, the highest sensitivity available among deterministic CCIT methods.

The primary application is where pressure and electrical methods reach their limits: lyophilized vials, dry powder formats, and cell and gene therapy products stored at -80°C or below. At cryogenic temperatures, elastomeric closures stiffen and defect morphology shifts in ways that reduce the reproducibility of ambient-condition pressure tests.

The method requires helium-filled headspace, making some configurations semi-destructive. For products where the sensitivity requirement justifies it, no currently available deterministic method offers comparable detection limits.

Conclusion

The shift from probabilistic to deterministic CCIT is both a regulatory expectation and an operational necessity. Batch destruction costs, the inability to generate quantitative defect data for regulatory submissions, and the sensitivity limitations of visual inspection are unsustainable for the biologics-dominated pipeline.

As per FDA guidance on container closure systems for packaging human drugs and biologics, manufacturers must demonstrate their CCI approach provides adequate sensitivity for the container-closure system and product risk profile. Vacuum decay, HVLD, and helium leak detection each provide a validated, quantitative pathway to that standard, but the right choice depends entirely on package format, fill matrix, and required detection threshold.

Frequently Asked Questions

1. What is the difference between probabilistic and deterministic CCIT methods?

Probabilistic methods such as blue dye ingress produce qualitative pass/fail results and destroy the test sample. Deterministic methods use physical measurement to generate quantitative leak rate data without compromising the package. USP <1207> explicitly prefers deterministic methods where technically feasible.

2. Is container closure integrity testing required by the FDA?

The FDA does not mandate a specific test method, but its guidance on container closure systems expects manufacturers to demonstrate their CCI approach is appropriate for the product and container format. For sterile drug products, this effectively requires a validated, sensitivity-justified CCI program.

3. Which CCIT method is best for vials and parenteral packaging?

The optimal CCIT method depends on the product, package format, and application requirements. For liquid-filled parenteral vials, High Voltage Leak Detection (HVLD) supports high-throughput deterministic inspection and is well suited for inline production environments. Vacuum Decay (ASTM F2338) is commonly applied to lyophilized products and headspace-containing packages where non-destructive leak detection is required. For cryogenic and ultra-high sensitivity applications, helium leak detection provides quantitative leak rate measurement with sensitivity beyond many conventional deterministic methods.

4. What does USP <1207> require for CCIT method validation?

Validation must demonstrate reliable defect detection at the sensitivity level appropriate for the product and container. Required studies include method suitability, specificity, detection limit, and robustness — each conducted for the specific container-closure system, not the method in isolation.

5. Can the same CCIT method be used across different container formats?

Not without separate validation. A vacuum decay method validated for 2 mL vials cannot be transferred to 50 mL bottles without re-establishing test parameters. Package geometry, headspace volume, and material properties each independently affect test performance.

Readmore...
: CCIT, container closure integrity testing, USP 1207, package integrity testing, pharmaceutical packaging, helium leak detection, vacuum decay, ASTM F2338, HVLD
584

Popular Blogs

Tags

GMP Annex 1 - How the Revision Impacts Pharmaceutical Manufacturers

Aug 21, 2023   |   18425

Stay ahead in pharmaceutical manufacturing excellence! Explore the transformative changes brought by the revised Annex 1 of EU GMP guidelines. Enhance your understanding of contamination control, quality risk management, and sterile manufacturing techniques. Learn how to implement these changes effectively and ensure product quality, safety, and efficacy.

Decoding USP Standards - An In-Depth Look at USP <382>

Aug 02, 2024   |   16775

USP <382> sets new standards for assessing the functionality of elastomeric components in injectable drug packaging, emphasizing comprehensive container closure integrity testing to ensure safety and efficacy. Compliance by December 2025 is critical for pharmaceutical companies to maintain robust and reliable packaging systems.

How to Leak Test High Risk Medical Device Packages

Jan 25, 2022   |   15213

Medical device leak testing is considered a critical step in ensuring package integrity primarily due to its sensitive nature. Leaks in the device can cause serious health risks to the patients and result in huge financial losses to the manufacturer. PTI's Seal-Scan technology is a non-destructive seal quality inspection technique for high risk pharmaceuticals and medical devices.

A Closer Look at PTI's Advanced Testing Services

Jun 27, 2024   |   12846

PTI offers advanced lab testing services across industries, safeguarding product integrity, safety, and regulatory adherence. Their expertise spans feasibility studies, test method development, stability testing, recall & batch release, and helium testing, all conducted by qualified professionals using advanced technologies.

Understanding Pressure Decay Leak Testing

Jun 07, 2021   |   12674

Pressure Decay leak test is a sensitive leak detection method particulary for testing small leaks. It is applicable across multiple industries like pharmaceutical, medical device and food and nutrition industries.
ptiusa

Our technologies conform to ASTM and other regulatory standards.

Packaging Technologies & Inspection

PTI offers inspection systems for package leak testing, seal integrity and container closure integrity testing (CCIT). Our technologies exclude subjectivity from package testing, and use test methods that conform to ASTM standards. PTI's inspection technologies are deterministic test methods that produce quantitative test result data. We specialize in offering the entire solution including test method development and equipment validation.

Sales Channel Partner Portal Login

Get in Touch

Packaging Technologies & Inspection

PTI offers inspection systems for package leak testing, seal integrity and container closure integrity testing (CCIT). Our technologies exclude subjectivity from package testing, and use test methods that conform to ASTM standards. PTI's inspection technologies are deterministic test methods that produce quantitative test result data. We specialize in offering the entire solution including test method development and equipment validation.

Sales Channel Partner Portal Login

ptiusa

Our technologies conform to ASTM and other regulatory standards.

Get in Touch

Popup