Blogs

29
Mar 2022

E-Scan MicroCurrent Technology - Quality Assurance Solution for Parenteral Product Packaging

E-Scan MicroCurrent Technology - Quality Assurance Solution for Parenteral Product Packaging

As technological advancements lead to an ever-increasing world of routes of administration for new and existing drugs, packaging decisions for such options have become more challenging. There are multiple options for drug delivery container formats, and each should be continuously reviewed with reference to compliance and accuracy of delivery. Packaging is a critical point of concern whenever a new drug product is introduced into the market. When it comes to parenterals, there has been a dramatic increase in these packaging formats over the past 10 years. Apart from typical formats of vial and syringes, dual-chamber devices, cartridges and electronically enabled devices have been introduced, all which demand high levels of packaging accuracy. So how do we ensure the ability of the packages to maintain sterility of the drug? Quality assurance with the proper Container Closure Integrity Test (CCIT) method is critical.

CCI Testing using E-Scan 655 MicroCurrent HVLD

The E-Scan 655 technology utilizes the MicroCurrent conductivity test method to non-destructive evaluate container closure integrity. MicroCurrent technology exposes the package and product to lower voltage than other conductivity based solutions. This unique technology requires no sample preparation and is a non-contact and non-invasive test method. What makes E-Scan 655 technology unique is its ability to test a wide range of liquid-based products including low conductivity sterile water for injection (WFI) and proteinaceous products with suspensions. The system also features a fast test cycle and is simple to operate. Additional benefits include quick product changeover and an easy recipe set up to accommodate a wide range of products and applications. The offline E-Scan 655 method can be migrated from the laboratory to automated 100% inline testing applications at high production speeds.

Testing Procedure

Using a set of electrode probes, E-Scan system tests a non-conductive container that is sealed. The container material can vary from glass, plastic, or poly laminate. The container or package must contain liquid (minimum fill 30%). In case of any defect including pinhole or crack, there will be a resistance differential and change in current flow indicating a breach in the container. The approximate defect location can be identified.

Benefits of MicroCurrent HVLD technology:

  • Non-destructive, non-invasive, no sample preparation
  • High level of repeatability and accuracy
  • Effective across all parenteral products, including extremely low conductivity liquids (WFI)
  • Lower voltage exposure produces no ozone, eliminating risk to the product and environment
  • Listed in USP Chapter <1207> as recommended method for parenteral liquid package integrity testing
  • Robust method and approximate 3x Signal-Noise-Ratio for a wide range of product classes and package formats
  • Simplifies the inspection and validation process
Readmore...
E-Scan, microCurrent hvld, parenteral product packaging, ccit, container closure integrity
937

Popular Blogs

Tags

How to Leak Test High Risk Medical Device Packages

Jan 25, 2022   |   7809

Medical device leak testing is considered a critical step in ensuring package integrity primarily due to its sensitive nature. Leaks in the device can cause serious health risks to the patients and result in huge financial losses to the manufacturer. PTI's Seal-Scan technology is a non-destructive seal quality inspection technique for high risk pharmaceuticals and medical devices.

CCI Inspection of Vials with Helium Leak Detection Technology

Jul 12, 2023   |   4658

Helium leak detection technology has revolutionized container closure integrity testing in the pharmaceutical industry by providing highly sensitive and precise detection of leaks in vials. By utilizing the unique properties of helium gas, this method enhances the overall integrity of vial packaging, ensuring the safety and quality of pharmaceutical products.

Container Closure Integrity Testing (CCIT) - An Ideal Solution for the Pharmaceutical Industry

Sep 01, 2020   |   4392

Container Closure Integrity Testing is a method of leak detection using a non-destructive packaging inspection system to prevent possible contamination in pharmaceutical products.

Understanding Pressure Decay Leak Testing

Jun 07, 2021   |   3615

Pressure Decay leak test is a sensitive leak detection method particulary for testing small leaks. It is applicable across multiple industries like pharmaceutical, medical device and food and nutrition industries.
ptiusa

Our technologies conform to ASTM and other regulatory standards.

Packaging Technologies & Inspection

PTI offers inspection systems for package leak testing, seal integrity and container closure integrity testing (CCIT). Our technologies exclude subjectivity from package testing, and use test methods that conform to ASTM standards. PTI's inspection technologies are deterministic test methods that produce quantitative test result data. We specialize in offering the entire solution including test method development and equipment validation.

Sales Channel Partner Portal Login

Get in Touch

 

Packaging Technologies & Inspection

PTI offers inspection systems for package leak testing, seal integrity and container closure integrity testing (CCIT). Our technologies exclude subjectivity from package testing, and use test methods that conform to ASTM standards. PTI's inspection technologies are deterministic test methods that produce quantitative test result data. We specialize in offering the entire solution including test method development and equipment validation.

Sales Channel Partner Portal Login

ptiusa

Our technologies conform to ASTM and other regulatory standards.

Get in Touch

 
Popup